End to End Solutions
Learn how Abbott’s End to End Solution unlocks synergistic value, helping your lab achieve and sustain excellence.
The ARCHITECT i4000SR, a multi-module system consisting of two ARCHITECT i2000SR systems, shares all the proven benefits of the ARCHITECT family of instruments. The ARCHITECT i4000SR provides enhanced productivity for labs with high-volume immunoassay demands. A large load-up capacity allows the ARCHITECT i4000SR to deliver high throughput for high-quality STAT results when you need them.
The ARCHITECT i4000SR immunoassay analyzer offers a maximum throughput of up to 400 tests per hour. Featuring a load-up capacity of 285 samples with 35 priority and 250 routine areas, the ARCHITECT i4000SR has 50 refrigerated reagent positions.
For in vitro diagnostic use only.
Description | |
Methods | CHEMIFLEX |
Maximum Throughput | Up to 100 tests/hour |
Sample Types | Serum, Plasma, Whole Blood, Urine |
Sample Tubes | Height: 72-102 mm Diameter: 9.6-16.1 mm |
Sample Cup | Yes (50 µl dead volume) |
Sample Capacity | 65 |
Sample Barcode Types | Code 39, Codabar, Interleaved 2 of 5, Code 128 |
Sample Result Storage | 50,000 |
Sample Volume | 10-150 μL, Average: 62 µL |
Automatic Dilution | Yes |
Sample Probe Carryover | ≤0.1 parts per million |
Reagent Capacity | 25 refrigerated positions |
Reagent Type | 100% liquid ready-to-use |
Reagent Onboard Stability | 14-30 days |
Calibration Frequency | Calibrate with new lot number, if controls are out of range or if specified otherwise within the package insert |
Sample, Clot and Bubble Detection | Yes |
Reagent Pressure Monitoring | Yes |
Sample Interference Measurement | No |
System Control Center | 1 SCC, with color touchscreen monitor, keyboard and mouse |
Onboard Maintenance Records | Yes |
Online Error Code Help | Yes |
Host Interface | Bidirectional, serial RS-232 interface, host query option available |
Remote Diagnostics | AbbottLink |
Dimensions | 49” x 59” x 30” 124.5 x 149.9 x 76.2 cm |
Weight | 636 lbs 288 kg |
Electrical Requirements | AC 110-120V or 200-240V, ±10%, 50 or 60 Hz self-adjusting |
Water Requirements | Purified water to dilute buffer concentrate |
Heat Output* | 2400 BTU/hr, running mode |
Sample Loading | RSH |
*Values provided represent the typical output in “Running” mode for the system processing module and sample handler. A maximum value of 389 BTU/hr was obtained for the System Control Center (SCC) using the ARCHITECT i2000SR as a representative system. Values obtained during an internal study. Data is on file with Abbott. |
Review a few details about the ARCHITECT i1000SR immunoassay analyzer in the specifications sheet (PDF, 118 KB).
Diagnostics at Abbott offers a complete portfolio of integrated solutions including Core Laboratory, Molecular, Point of Care, Rapid Diagnostics and Transfusion Medicine.
Learn how Abbott’s End to End Solution unlocks synergistic value, helping your lab achieve and sustain excellence.
Find the latest information in laboratory science, therapeutic areas and Abbott innovation.
Leverage The Lab
Learn how Abbott's strategic transformation plan provided laboratories insights and health system performance improvements.
Links which take you out of Abbott worldwide websites are not under the control of Abbott, and Abbott is not responsible for the contents of any such site or any further links from such site. Abbott is providing these links to you only as a convenience, and the inclusion of any link does not imply endorsement of the linked site by Abbott. The website that you have requested also may not be optimised for your screen size.
Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions.